Optical transmittance of silica aerogel

Additional data

Submitted: 28.11.2025; Accepted: 20.02.2026; Published 04.03.2026;
Views: 10; Downloaded: 6

How to Cite

A. Hakobyan, A. Mkrtchyan, D. Khurshudyan. "Optical transmittance of silica aerogel" Natural Sci. Rev. 3 100602 (2026)
https://doi.org/10.54546/NaturalSciRev.100602
A. Hakobyan1,a, A. Mkrtchyan1, D. Khurshudyan1
  • 1Alikhanyan National Science Laboratory, Yerevan, Armenia
  • aargine.hakobyan@yerphi.am
DOI: 10.54546/NaturalSciRev.100602
Keywords: aerogel, SPD, NICA, detector, transmittance measurement
Topics: Physics , High Energy Physics (Experiment)
PDF

Abstract

This paper presents the investigation of aerogel Cherenkov radiation detectors and the optical characterization of aerogel samples within the framework of the SPD experiment at the NICA accelerator complex, which is under construction at the Joint Institute for Nuclear Research (Dubna, Russia). The transmittance of aerogel samples was measured at the A. Alikhanyan National Science Laboratory (AANL) (Yerevan Physics Institute). Longitudinal transmittance measurements were performed for two distinct aerogel samples. Data analysis and visualization of results were carried out using the OriginPro 8.5 and ROOT software packages. The obtained results were compared with analogous studies performed at other research centers in order to evaluate the reliability, consistency and accuracy of the measurements.

Acknowledgements

We thank H. Mkrtchyan and V. Tadevosyan for assistance and technical support at various stages of the work. We are especially grateful to A. Barnyakov for provided samples of aerogel material produced at BINP.

References

[1] V. Abazov et al., Technical Design Report of the Spin Physics Detector at NICA, Natural Science Review 1 (2024) 1. arXiv:2404.08317.

[2] M. Cantin, M. Casse, L. Koch, R. Jouan, P. Mestreau, D. Roussel, C. Saclay, F. Bonnin, J. Moutel, and S. J. Teichner, Silica aerogels used as Cherenkov radiators, Nuclear Instruments and Methods 118 (1974) 177–182.

[3] T. Sumiyoshi et al., Silica aerogel Cherenkov counter for the KEK B-factory experiment, Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 433 (1999) 385–391.

[4] V. V. Anashin et al., Physics of Elementary Particles and Atomic Nuclei 44 (2013) 1263 (in Russian).

[5] I. V. Ovtin et al., Aerogel Cherenkov Counters of the KEDR Detector, CERN–BINP Workshop for Young Scientists in e+e− Colliders, 22–25 Aug. 2016, CERN-Proceedings-2017-001, CERN, Geneva, 2017.

[6] T. Iijima et al., Aerogel Cherenkov counter for the BELLE detector, Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 453 (2000) 321–325.

[7] A. Abashian et al., The Belle Detector, Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 479 (2002) 117–232.

[8] Y. Allkofer, C. Amsler, S. Horikawa, C. Regenfus, and J. Rochet, A new aerogel Cherenkov detector for DIRAC-II, Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 595 (2008) 84–87.

[9] M. Contalbrigo, G. Adams, P. Bosted, et al., The aerogel Cherenkov detector for the SHMS magnetic spectrometer in Hall C at Jefferson Lab, Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 805 (2016) 12–16. https://www.sciencedirect.com/science/article/pii/S0168900216310774.